Health product recall

3m™ Steri-Drape™ Surgical Drapes

Brand(s)
Last updated

Summary

Product
3m™ Steri-Drape™ Surgical Drapes
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

3m™ Steri-Drape™ Surgical Drapes

More than 10 numbers, contact manufacturer.

1020NS
1030NS
1033
1060NS
1071
1021NS
1030
1000NS
1021
1060
1061
1010
1010NS
1020
1061NS
1000

Issue

During a recent investigation, 3m confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Although the acceptance criteria for the device was met upon product release, the adhesive component may not meet 3m's performance criteria throughout product shelf life. Additionally, 3m has observed an increase in reported adhesive related skin injuries for these affected lots.

Recall start date: October 7, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

3m™ Steri-Drape™ Surgical Drapes

2510 Conway Avenue, St. Paul, Minnesota, United States, 55144

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72051

Get notified

Receive notifications for new and updated recalls and alerts by category.